Long-term: Stored in a buffer solution of pH=7.
2. Cleaning of pH glass electrode The contamination of the glass electrode bulb may lengthen the electrode response time. The soil can be decontaminated with CCl4 or soap, and then immersed in distilled water for a day and night to continue use. When the pollution is serious, immerse in 5% HF solution for 10-20 minutes, rinse immediately with water, then immerse in 0.1N HCl solution for a day and night and continue to use.
3. Treatment of glass electrode aging The aging of the glass electrode is related to the gradual change of the rubber layer structure. The old electrode has a slow response, high membrane resistance, and low slope. Etching the outer layer with hydrofluoric acid often improves electrode performance. If this method is used to periodically remove the inner and outer layers, the life of the electrode is almost unlimited.
4. The best storage solution for the silver-silver chloride electrode of the reference electrode is a saturated potassium chloride solution. The high concentration potassium chloride solution can prevent the precipitation of silver chloride at the liquid junction and maintain the liquid junction at work. status. This method is also applicable to the storage of composite electrodes.
5. Regeneration of reference electrode Most of the problems caused by the reference electrode are caused by blockage of the liquid junction, which can be solved by the following methods:
(1) Soaking liquid junction: using a mixture of 10% saturated potassium chloride solution and 90% distilled water, heating to 60-70 ° C, immersing the electrode in about 5 cm, and soaking for 20 minutes to 1 hour. This method dissolves the crystals at the ends of the electrodes.
(2) Ammonia soaking: When the liquid junction is blocked by silver chloride, it can be diluted with concentrated ammonia. The specific method is to wash the electrode inside, and immerse the liquid in ammonia water for 10 to 20 minutes, but do not let ammonia water enter the inside of the electrode. The electrode was taken out and washed with distilled water, and the internal liquid was refilled and continued to be used.
(3) Vacuum method: The hose is placed around the reference electrode liquid junction, and the water suction pump is used, and the liquid in the suction part passes through the liquid junction to remove the mechanical blockage.
(4) Boiling solution junction: The liquid junction of the silver-silver chloride reference electrode is immersed in boiling water for 10 to 20 seconds. Note that the electrode should be cooled to room temperature before the next boil.
(5) When the above methods are invalid, the mechanical method of sandpaper grinding can be used to remove the blockage. This method may cause the sand under grinding to be inserted into the liquid junction. Causes permanent blockage.
(2) Inspection 1. General inspection method of glass electrode (1) Check the zero potential setting pH meter in the “mV†measurement file, insert the glass electrode and the reference electrode into the buffer solution with pH=6.86. The reading of the instrument should be about It is -50 to 50 mV.
(2) Check the slope Connect to (1), then measure the mV value of the buffer solution with pH=4.00 or pH=9.18, calculate the slope of the electrode, and the relative slope of the electrode should generally be composite technical index.
note:
1) The electrode zero potential value check method is only for the glass electrode with an isoelectric point of 7. If the equipotential point of the glass electrode is not 7, it is different.
2) For some pH meters, when the calibration adjustment can meet the requirements, the electrode can be used when the above inspection result is out of range.
3) For some intelligent pH meters, you can directly check the zero potential and slope values ​​obtained from the calibration results of the instrument.
2. Reference electrode inspection method (1) The internal resistance inspection method uses a laboratory conductivity meter. One end of the conductivity socket of the conductivity meter is connected to the reference electrode, and the other end is connected with a wire. The reference electrode and the wire are simultaneously immersed. In the solution, the measured internal resistance should be less than 10kΩ. If the internal resistance is too large, it indicates that the liquid junction is blocked and should be treated.
(2) Electrode potential check: Take a good reference electrode of the same model and the reference electrode to be measured into the input end of the pH meter, and then insert the KCl solution (or buffer solution with pH=4.00). The potential difference should be -3 to 3 mV, and the potential change should be less than ±1 mV. Otherwise, the reference electrode should be replaced or regenerated.
(3) Appearance inspection Silver-silver chloride wire should be dark brown. If it is grayish white, it indicates that silver chloride has partially dissolved.
Classification of testing reagents on the market:
1. Nucleic acid detection reagents((2019-nCoV) Nucleic Acid Diagnostic Kit) are medical reagents, with high accuracy and high price. The throat takes liquid samples for testing. It takes 2 hours to produce the results. It requires professional equipment. The professionalism is high. The compliance rate is 97%. Degrees of preservation, dry ice transportation! Suitable for hospitals, government agencies!
2. Antibody Detection Reagents(COVID-19 IgM/IgGAntibody Detection Kit) are civilian detection reagents that need instrumental assistance. The accuracy is average. The blood test can produce results in 15-30 minutes. It is fast and simple, and the price is low (about 8.5 US dollars). It is shipped at room temperature! Suitable for civilian use
3. Colloidal gold,test kits(Colloidal Gold COVID-19 IgM/IgG Detection Kit), civil reagents, blood tests, pregnancy test papers, no instruments required, fast results in 5 minutes, a few minutes, 8 US dollars each! Normal temperature transportation.
Nucleic Acid Diagnostic Kit
Pcr Instrument,Antibody Detection Kit,Antibody Detection Reagents,Nucleic Acid Diagnostic Kit
Guangzhou Zhongzhinan Supply Chain Co.,Ltd. , https://www.gzzhongzhinan.com