Direct Bilirubin (D-Bil) Assay Kit/Test Kit/Ivd Reagents

Model NO.: D-Bil
Application: Hospital, Lab, School
Trademark: Biomedia
Specification: R1: 2*60 mL, R2: 2*15 mL
Origin: Jinan, Shandong
Model NO.: D-Bil
Application: Hospital, Lab, School
Trademark: Biomedia
Specification: R1: 2*60 mL, R2: 2*15 mL
Origin: Jinan, Shandong


The direct bilirubin (D-Bil) assay kit provides a convenient and accurate measurement of concentration of D-Bil, also called conjugated bilirubin, in human plasma.

Test principle

In the presence of surfactant and indirect bilirubin inhibitors, bilirubin can be oxidized by sodium nitrate and converted into biliverdin, which leads to a decrease in absorbance at 450 nm. Total bilirubin concentrations can be calculated from the absorbance change measured at 450 nm.

Product Features
 
1. Linear range of D-Bil level: [0, 342] μmol/L
Linear correlation coefficient ∣r∣≥0.990
Linear deviation ≤ ± 10.00%
2. Absorbance of blank ≤ 0.10  
3. Sensitivity
The change in absorbance (ΔmA) ≥ 0.30 while testing the specific D-Bil sample of concentration of 47 μmol/L.
4. Accuracy
Relative deviation is within ±10%.
5. Precision
Intra CV% ≤ 5.0%; inter CV% ≤5.0%.

Technical Parameters
 
Temperature 37ºC Optical path length (OPL) 1.0 cm
Main wavelength 450 nm Secondary wavelength 570 nm (Optional)
Method End-point method
(dual reagents)
Reaction direction Negative
Sample 8 μL R1 240 μL
R2 60 μL Reaction time 300 s
Reference range: 0.40 - 6.64 μmol/L

Packing Specifications

R1: 2×60 ml, R2: 2×15 ml
R1: 4×60 ml, R2: 4×15 ml
R1: 1×40 ml, R2: 1×10 ml (Sample)

Storage and stability

Unopened: Store assay kits at 2 - 8ºC protected from light. Stable to expiration date printed on individual vials (for 16 months).
Opened: Store assay kits at 2 - 8ºC protected from light. Stable for 1 month if kept capped from original container and free from contamination.
Do not freeze the assay kit.
 

The direct bilirubin (D-Bil) assay kit provides a convenient and accurate measurement of concentration of D-Bil, also called conjugated bilirubin, in human plasma.

Test principle

In the presence of surfactant and indirect bilirubin inhibitors, bilirubin can be oxidized by sodium nitrate and converted into biliverdin, which leads to a decrease in absorbance at 450 nm. Total bilirubin concentrations can be calculated from the absorbance change measured at 450 nm.

Product Features
 
1. Linear range of D-Bil level: [0, 342] μmol/L
Linear correlation coefficient ∣r∣≥0.990
Linear deviation ≤ ± 10.00%
2. Absorbance of blank ≤ 0.10  
3. Sensitivity
The change in absorbance (ΔmA) ≥ 0.30 while testing the specific D-Bil sample of concentration of 47 μmol/L.
4. Accuracy
Relative deviation is within ±10%.
5. Precision
Intra CV% ≤ 5.0%; inter CV% ≤5.0%.

Technical Parameters
 
Temperature 37ºC Optical path length (OPL) 1.0 cm
Main wavelength 450 nm Secondary wavelength 570 nm (Optional)
Method End-point method
(dual reagents)
Reaction direction Negative
Sample 8 μL R1 240 μL
R2 60 μL Reaction time 300 s
Reference range: 0.40 - 6.64 μmol/L

Packing Specifications

R1: 2×60 ml, R2: 2×15 ml
R1: 4×60 ml, R2: 4×15 ml
R1: 1×40 ml, R2: 1×10 ml (Sample)

Storage and stability

Unopened: Store assay kits at 2 - 8ºC protected from light. Stable to expiration date printed on individual vials (for 16 months).
Opened: Store assay kits at 2 - 8ºC protected from light. Stable for 1 month if kept capped from original container and free from contamination.
Do not freeze the assay kit.
 