GY-4 digital fruit hardness tester instruction manual

GY-4 digital fruit hardness tester instruction manual
Thank you for purchasing the GY-4 series digital display fruit hardness tester:
The fruit hardness tester has the advantages of high precision, easy operation, and convenient carrying. More test modes are available for testers to use, and test data can be stored and input to a computer for statistical analysis. It can be used with a dedicated test bench to improve measurement accuracy.
First, the functional characteristics:
1. High-precision and high-resolution: large-screen LCD display with backlight function (backlight suitable for night use), and with screen digital positive and reverse reversal functions;
2, with battery capacity display: divided into 3 grids, 2 grids, 1 grid, the instrument automatically shuts down when the power is too low;
3, with gravity acceleration setting function: the user can input the exact value of the gravity acceleration of the place of use, so that the test is more accurate;
4, with peak hold function: keep the peak display until manually cleared;
5, with automatic peak function: keep the display peak automatically after 2 seconds;
6, with a large memory storage function: can store 896 test values;
7, with data output function: the data can be input into the computer through the data line for various analysis;
8, with automatic shutdown time setting: automatic shutdown time setting can be set to automatically shut down from 10 minutes to 90 minutes;
9, with high-quality charging power: charging voltage from 100V to 240V available, can adapt to most areas of China and abroad;
10, with 2 sets of installation dimensions: adapt to most test machines at home and abroad, convenient for users to install on the machine.
Second, technical parameters:
1, model: GY-4;
2. Unit: Kg/cm2 (×105 Pa);
3. Indenter diameter: 1 # :Φ11.1mm, 2 # :Φ7.9mm
4. Range: 1 # : 0.2~15 Kg/cm 2 (×10 5 Pa), 2 # : 0.4~30 Kg/cm 2 (×10 5 Pa);
5. Graduation value: 0.01 Kg/cm 2
6. Indication error: ±0.5%;
7, accuracy: ± 1%;
8, the indentation depth: 10mm;
9, power: charging power: 220V / AC; battery continuous working time: 6 ~ 8 hours;
10. Power supply mode: No. 5 nickel cyanide battery pack / 220V AC charging 4 to 6 hours
11, stability: temperature drift: 0.2uV / °C (0-60 ° C); zero drift: ≤ 0.1% / 8 hours / FS;
12, working temperature: 5 ° C ~ 35 ° C
13, the relative temperature: 0 ~ +60 ° C;
14, ambient humidity: 15% ~ 80% RH;
15. Working environment: no source and corrosive medium around;
16, net weight: ~ 450g;
17. Dimensions: 230×66×36mm.
Third, the structure diagram:
Fourth, the button description (such as the combination of the next two figures):
1 , "boot" button:
When this button is pressed, the power is turned on and the indicator light will be on, and the model number will be displayed on the LCD. When the LCD shows 0.00, it can be operated. (Note: Sometimes after turning on the power, there may be a zero-drift value display of the analog signal on the screen, that is, the value display drifts up and down at zero point. Clear the “clear” button to clear it;
2 , "Shutdown" button:
Press this button to turn off the power. (Note: Whenever you press this button, the power is turned off, but the stored memory data will not disappear);
3 , "memory" button:
When this button is pressed, the stored test data (memory data) will be recalled and displayed on the screen in turn. At this time, the word “MEM” flashes first to display the number of memories, and after 2 seconds, the memory data is automatically displayed. If you want to exit the memory function interface, just press the “clear” button to exit;
4 , "Unit" button:
Press this button to switch the measurement unit. Provide N (Newton), Kg (kg), and lb (lb). Each time you press this button, the unit will switch to the next unit.
5 , "peak" button:
Press this button to switch between manual peak hold, automatic peak hold, and load real-time value. “PEAK” displayed on the LCD indicates manual peak hold, “AUTO PEAK” indicates automatic peak hold, no “PEAK”. "and "AUTO PEAK" display shows the real-time value of the load. Each time the button is pressed, the mode switches to the next mode;
6 , "storage" button:
When this button is pressed, the test data displayed on the screen is stored (in the manual peak state). The stored data can be called up by the “memory” button, cleared by the “clear” button, or output to the computer for analysis or printing. When there is test data stored in the machine, there will be a “MEM” display on the LCD.
7 , "clear" button:
When you press this button, the test value on the screen will be reset to zero.
※ When the weight of the clamp used exceeds 20% of the range, press the “clear” button to clear it. In this case, you need to select the lighter fixture first, and then press the “clear” button to clear it.
※ When the load used exceeds 20% of the range of the machine, press the “clear” button to clear it. In this case, the load must be released first, and then press the “clear” button to clear it.
※ Press and hold this button for 4 seconds, the stored test data can be completely cleared (Note: In some states, it may not be cleared, you need to turn off the power first, turn it off and then turn it back on, then press this function key to clear all memory data. );
8 , "communication" button:
Press this button to switch the screen number positive and reverse.
Five, the screen display instructions
1 , PEAK
When “PEAK” is displayed, it indicates Peak mode, the display shows peak value until manual clear; when “AUTO PEAK” is displayed, it means “auto peak” (peak is automatically released mode), and the peak value is displayed for 2 seconds. After the automatic clear; when "PEAK" is not displayed, it means "track mode", the value on the screen changes with the load;
2 , LO BAT
When the battery voltage drops below 7.0V, “LO BAT” appears on the screen, indicating that the voltage is insufficient and needs to be recharged (it can still be tested while charging);
3 , MEM
When there is data stored in memory, “MEM” will be displayed. When the “DATA” button is pressed to view the memory data, “MEM” flashes.
6. Storage (memory) information:
1. Store data
In the peak hold (PEAK) state, after the test is completed, press the “AUTO PEAK” button to save the data. The display will show “MEM” and the stored memory data can be saved after shutdown. When you use the “Memory” button to view the memory data, “MEM” flashes, the number of memories first appears, and the memory value appears automatically after 2 seconds. Press the “Peak” button to exit the memory and enter the test directly. Data can also be entered into the computer for analysis. This unit can store 896 data. When "OE" is displayed, it means that the 896th data can no longer be stored.
2 , memory clearing
In the general load display state, press the "clear" button for more than four seconds, all the memory data will be cleared, and the word "MEM" disappears.
This unit can be connected to a computer to input test data to the computer. View, print the number of tests, average, maximum, minimum, and determine whether the test results meet the set requirements.
Seven, other function description:
1 , print function description:
This function prints the stored test data to the computer through the data line.
A. First run the CD-ROM that is carried randomly (or download the “Digital Display Force Synchronization Test Software Download” in the problem column on the homepage of the company → “http://” After that, you will get the folder, open it and click the "Data Export Program" file, then click the "setup.exe" file, then click "Next" → "Next" → "Complete".
B. Connect the instrument to the computer with the data cable. When the instrument is powered on, when the LCD shows zero, double-click the “Densometer Communication Software 3.0.exe” shortcut on the desktop, and then pop up. Click "Receive" in the window, and the stored test data is output to the computer through the data line. After the data is received, click “Print” and click Print in the pop-up window to print.
2 , synchronization test function description:
This function displays functions such as synchronized test curves and data during testing.
A. First run the CD-ROM that is carried randomly (or download the “Digital Display Force Synchronization Test Software Download” in the problem column on the homepage of the company → “http://” After that, you will get the folder, open it and click the “Sync Test Software” file, then click “Install.EXE” file, then click “Next” → click “Next” → “Complete”.
B. For specific operation steps, please download the “Digital Display Force Synchronization Test Function Demonstration” in the problem column on the homepage of our company, and refer to the operation of this demonstration.
3 , to solve the overload status description:
When the power-on alarm and the display data cannot be cleared, the battery is low or severely overloaded.
Approach:
A, charging.
B. Press and hold the “Settings” button for more than 4 seconds, it will automatically shut down. After turning off the power, press the “Power On” button to turn it on and restore the factory settings (if it cannot be recovered, you need to return to the factory to replace the sensor).
4 , to solve the crash status description:
When the measuring instrument freezes, press the “Reset” button on the left side of the measuring instrument.
Eight, the change of gravity acceleration value
The user of this machine can input the gravity acceleration value of the place of use. Press and hold the “Peak” button before powering on, and then press the “Power On” button to enter the gravity acceleration value setting interface. After inputting the new gravity acceleration value, press the “Set” button to shut down the system. Reboot to use.
Nine, correction method
1. Synchronous test software can be connected to the computer test. The test force curve graph and the detailed test force record during the test can be displayed synchronously on the computer, and can be saved, printed, and analyzed. (For details, see CD-ROM information)
2. Automatically store test value function In the “AUTO PEAK” state, the instrument will automatically store the test peak value. Press the “Memory” button to scroll through the stored data, or import the data through the “Data Export Program” on the companion CD. Computer, the program will automatically display the "maximum value, minimum value, average value, deviation value, qualified maximum value, qualified minimum value, qualified average value, test number, pass number" and other data for users to analyze and use (see CD-ROM for details) Corresponding instructions) When connecting with the computer to export data, the interface is friendly, connected through the matching data cable, and can be used under the normal boot interface. Provide both Chinese and English interface programs;
3. The extra large memory storage function can store 896 test values. It can be stored manually in the state of “PEAK” (press the “Peak” button once) or automatically in the state of “AUTO PEAK” (press the “Peak” button twice). Press the “Memory” button to view the stored memory values ​​in turn. You can also view the stored memory values ​​in turn by pressing the “Memory Export” button on the companion CD, or by using the “Memory” button on the companion CD. ", import data into the computer;
4. Added backlight function. When the button or instrument is stressed, the backlight will automatically light up. If there is no action for 10 seconds, the backlight will be turned off automatically. You can also press the “Set” button 6 times to enter the “light” state and select “NO”. Turn off the backlight and select “YES” to turn on the auto backlight function. ;
5, LCD liquid crystal display flip display Press the "print" button to make the reading 180 degrees reverse display, convenient to view the test value from different angles;
6. Automatic peak function Press “Peak” button 2, “AUTO PEAK” is displayed on the screen, press “Save” button to enter the automatic peak setting interface, press “Print” button to set the automatic peak time (ie test value) The time remaining on the screen), from 1 second to 9 seconds, after the selected time, press the “Set” button to save and exit the automatic peak time setting. The default time when the instrument is shipped from the factory is 3 seconds;
7. The automatic shutdown time setting can be set to automatically shut down from 10 minutes to 90 minutes, or it can be set to not automatically shut down. Press the “Set” button 7 times to enter the “OFFT.” interface, press the “Print” button to select the automatic shutdown time, select “00” does not automatically shut down, press the “Set” button to save and exit. The instrument is set to 50 minutes by default;
8. The battery capacity display is divided into 3 grids, 2 grids and 1 grid. When the battery is too low, the instrument will automatically shut down;
9. Metering correction When the metering department needs to calibrate the instrument, press and hold the “memory” button while pressing the “on” button until the display of a row of numbers, press the “set” button to display “FULL”, then display A row of numbers, hang the corresponding weight of the full scale, press the “Set” button after stabilization, automatically shut down, the calibration ends, and the calibration data is stored in the instrument. If the correction value is incorrect, long press the “Set” button to return to the factory setting state; long press the “Save” button will return to the previous calibration state;
10, accidental crash fast reset on the side of the reset button, with a hard object press the reset switch, you can shut down.
X. Test procedure
1. The specific operation method of the instrument installed on the test machine:
A. Loosen the adjustment handle 1 first, and slide it down to contact the chassis 2 to expose the two mounting holes above the mounting plate 3. Then install the upper two holes of the instrument on the mounting plate 3 with mounting screws and spring washers. Finally, loosen the adjustment handle 1, and return it to the original upper position to lock it. (Figure 1 above)
B. After the handle 4 is pressed, the two mounting holes below the mounting plate 3 can be exposed; and the lower holes of the instrument are mounted on the mounting plate 3 by using mounting screws and spring pads. (Figure 2 above)
C. Follow the previous two steps to tighten the four screws.
2, select the appropriate probe (Note: the selection of the diameter of the probe 11, the instrument displays the hardness value hardness actual fruit; the selection of a diameter of the measuring head 8, the display device is two times the actual hardness value hardness of fruit Value ) , attach the probe to the test pole of the instrument with a connecting sleeve with internal thread. (Figure 3 above)
3. Before the measurement: Firstly, the fruit to be tested is peeled off by about 1 square centimeter. When measuring, the center of the fruit to be tested is in line with the test rod of the instrument, so that the force value of the test can be more accurate. (Figure 3 above)
4. Start measurement: press “Power On” to turn on the power. After the LCD display is stable, press “Peak” to enter the peak measurement mode (if the display is not zero, press “Zero” to clear), the handle will be 4 Press down to make the measuring head face the peeled pulp, evenly press it into the line, and complete the measurement. The displayed value is the hardness of the fruit. (Figure 4)
5. After the test is completed, remove the load, turn off the power, remove the fixture, and clean the objects and put them back into the tool box for the next use.
XI, safety precautions:
Precautions
If the operation is incorrect, it may damage the instrument or cause a serious accident. In this manual, important items for preventing accidents and how to use the instrument are pointed out. Please read this manual carefully before use and keep it for future reading.
Warning
1. Do not use the instrument beyond the maximum range. Failure to do so may result in damage to the sensor or even an accident;
2. When the test value exceeds 105% of the range, the buzzer will continue to beep. In this case, please quickly release the added load or reduce the load.
Safety Precautions
1. Please use the matching charger to charge, otherwise it will cause circuit failure and even cause fire;
2. Do not use a power supply other than the rated voltage of the charger, otherwise it may cause electric shock or fire;
3. Do not pull out or insert the plug with wet hands, otherwise it may cause electric shock;
4. Do not pull the charger's power cord to pull out the plug, so as to avoid the electric wire being broken and being shocked;
5. Please clean the unit with a soft cloth. Dip the cloth into the water with detergent, wring it out and remove dust and dirt;
Note: Do not use volatile chemicals to clean the unit ( such as volatiles, thinners, alcohol, etc. )
6. Do not operate the unit in the following environments:
(1) a humid environment; (2) a dusty environment;
(3) Where oil or chemicals are used; (4) Where there is a source around.
7. Please use and store within the specified temperature and humidity range, otherwise it may cause instrument failure;
8. Do not disassemble or repair the machine by yourself. These actions may cause permanent malfunction of the instrument;
9. Other unsatisfactory matters needing attention in safety production.
Twelve, other matters needing attention
GY-4 type fruit hardness tester before delivery
power supply
symptom
Cause or phenomenon
Disposal
Press "Power On" button without display
Battery is dead
Recharge
Unable to charge
Charging using a converter that does not match the specifications
please confirm:
AC110V→DC9.4V AC220V→DC9.4V
test value
Inaccurate test value
Too much error
Weight correction reading (reference correction method)
other
Parameter confusion
MODE setting error
Press the “Set” button for more than 4 seconds to restore the factory settings.
Accidental crash
Press any key without reflection
Press the reset button and restart
Thirteen, shape and installation dimensions (The two sets of holes on the installation diagram are for fixing on a dedicated measuring frame, and the measuring frame is self-contained)
2 sets of installation dimensions, respectively: 4-M4—40×90
4-M3—30×145
Special Note:
GY series fruit hardness tester has a variety of specifications for users to choose, users can select the corresponding specifications according to the force value of the test product. The scientific test range is from 10% to 100% of full scale , and the metrology department does not recommend using less than 1% of full scale .

Band Aid

Band Aid,Waterproof Band Aids,Medical Health Band-Aids,Non-Woven Waterproof Band-Aids

Shaoxing Weijian Medical Supplies Co.,Ltd. , https://www.wetourmedical.com